Chapter A7: Set Preferences
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A7-48
Operator Manual © 2008 General Electric Company. All rights reserved.
Figure A7-31 Edit Techniques screen
19. Change the technique parameters as appropriate.
CAUTION: If no technical factors are present in the system, the default settings are:
•kV = 50
• mA = 100 (fixed setting)
•mAs = .05
No exposures should be made until the user selects values appropriate for the patient
size.
20. Upon completion for each view, press [NEXT TECH] adjust techniques for the next configured patient
size.
21. When finished, press [EXAM MENU].
The Exam Menu appears.
22. Press [SAVE].
A message appears: “Changes made to the Protocol Database will be saved.”
[RESET] clears all changes to the database since the last time it was saved.
23. Press [OK].
The Preferences – Protocols screen appears.