Chapter A8: Quality Assurance Process and Maintenance
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A8-4
Operator Manual © 2008 General Electric Company. All rights reserved.
Figure A8-4 Quality Assurance Program screen
3. Press [DETECTOR CHECK].
The Detector Check Tests screen appears.
Figure A8-5 Detector Check Tests screen
4. Press [START] to begin the Detector Check.
5. Wait for the test to complete.
The Detector Check result screen appears automatically after tests have been completed.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).