Chapter A1: Detector Overview
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A1-12
Operator Manual © 2008 General Electric Company. All rights reserved.
Ambient
•Temperature
Humidity
Pressure
External ambient temperature
range: +10 ºC to +30 ºC.
Rate of temperature change:
10 ºC per hour
Ambient humidity range: 10% to
80%, non condensing.
Rate of humidity change: 30% per
hour.
Atmospheric pressure (altitude)
range: -400 meters to +3000 meters
relative to sea level.
External ambient temperature
range: 0ºC to +50 ºC.
Rate of temperature change:
20 ºC per hour
Ambient humidity range: 5% to 95%,
non condensing.
Rate of humidity change: 30% per
hour.
Atmospheric pressure (altitude)
range: -100 meters to +15,000
meters relative to sea level (to
support pressurized transport)
Mechanical
Stress &
Vibration
Forces
The detector assembly shall not be
exposed to operating vibration
spectrum exceeding the following
parameters:
•Type: Random
Frequency Range: 20 to 350 Hz
Magnitude: 0.006g2 /Hz at 10-350
Hz
Duration: 8 hours/axis (x, y, z)
The detector assembly shall not be
exposed to non-operating vibration
spectrum exceeding the following
parameters:
•Type: Random
Frequency Range: 10 to 2000 Hz
Magnitude: 6 m/s2 RMS or
0.02g2 /Hz at 10-2000 Hz
Duration: 15 minutes/axis (x, y, z)
Shipping &
Storage
Environment
Not applicable. The non-operating shipping conditions
shall be -20 to +60 with the detector
and packing.
The shipping container shall protect
the detector from vibration of 2 Grms
for 8 hours in the x, y, and z axes,
random vibration from 10 to 2000 Hz
such that the image quality is not
degraded.
Cargo hold during shipment shall be
within the atmospheric pressure range
of 700-1100hPa