Section 4 Measurement Types and their Uses
103
4. Use EITHER of these methods, depending on the type of
measurement:
Press Spectrum Waveform > Spec Wfm if you are
taking a spectrum, waveform, demodulation, or coast-
down/run-up measurement.
Press Number of Samples if you are taking an Orbit Plot
measurement.
5. Use the arrow keys on the left to select the lines of resolution for the
measurement, or samples for orbit plots.
NOTE: Order tracked recordings are limited to 6400 lines of
resolution for single channel recordings, and 3200 lines for dual
channel recordings.
6. Press to save your changes and return to the main
measurement screen.
7. Use EITHER of these measurements, depending on the type of
measurement:
Press Fmax Fmin for measurements based on orders.
Press Wfm Duration Total Time for measurements based
on revs.
8. Press Ord Trk or Revs Trk or use the left and right arrow keys
to select the text Tracked from the select box > Press .
9. Enter the number of orders or revolutions that will be tracked.
NOTE: This value must be between 1 and 999.
NOTE: To revert the measurement to a non-order tracked
recording, press Ord or Revs > Enter a new value.
Alternatively, you can use the arrow keys to select an Fmax or the
text Orders or Revs > Press .

Table of Contents