Section 4 Measurement Types and their Uses
56
To take a 6Pack measurement:
1. Press or Measure.
2. Press 6Pack.
3. Use the arrow keys on the left to select a parameter set or press
to create a new set.
4. Set the parameters as required.
5. Press Spectrum Waveform, HF Fmax (Dmax), or
Demod Band (Dmin).
6. Use keys , , and to select the measurement types
you wish to record.
7. Press to begin sampling.
NOTE: The six recording types that comprise a 6Pack
measurement are saved as a single recording for each channel,
not as individual High Frequency, Low Frequency, and
Demodulation.
Streamlined Parameter Settings
The following simplifications have been applied to 6Pack measurements:
The sensor type must be Accelerometer.
The High Frequency and Demodulation measurements are
measured in Acceleration units only.
All spectra and waveforms have the same resolution (maximum
3200 lines, 40 kHz for single channel and 1600 lines, 20 kHz for
dual channel).
The Demod Bandwidth upper limit is automatically set to equal
the HF Fmax.
Averaging type, number, and overlap are the same for all
spectra.
Fmin and Demodulation Fmin are the same.
Fmax and Demod Fmax are the same.
NOTE: Several charts will display when you use the 6Pack feature,
especially when you measure with two channels.
MORE: See View Multiple Charts for more information on multi-
chart layout options.