Section 10 — Create, Edit, and Delete Items
208
To change the Bias Voltage range settings for a sensor:
1. Press Options on the main menu.
2. Press Sensor Setup.
3. Press Change Sensor beside the channel you wish to use.
4. Press Edit Sensor to edit the configuration of an existing
sensor, or Create New Sensor to create a new sensor.
5. Configure the sensor's normal parameters (name, type, settling,
units, etc).
6. Press Bias Voltage Check to open the Bias Voltage screen.
NOTE: Pressing toggles the Bias check on/off.
7. Press Create New Bias Voltage.
8. Press .
9. Enter a name > Press .
TIP: You can use the name of the sensor.
10. Press Min Bias > Enter the minimum bias Voltage level > Press
.
11. Press Max Bias > Enter the maximum bias voltage level > Press
.
12. Press twice to select the new bias check configuration — The
sensor will now use the new or edited bias values.
Set Default Sensor
When you specify a default sensor and enable its measuring channel,
your instrument will assume you are using this sensor to take all
subsequent measurements. You can change to a different sensor
temporarily or change the default sensor.