Presets and System setup
Vivid i User’s Manual 315
2378958-100 Rev. 02
To select the default dataflow
1. Select the dataflow in the Name drop-down menu. (see
Figure 12-13).
2. Check the Default box.
The dataflow will be selected by default when restarting
the unit.
3. Check the Direct Store box to have data stored
automatically to the archive (no buffer storage).
To create a new dataflow
1. Press Add.
The Add dataflow window is displayed.
Figure 12-14: The Add dataflow window
2. Enter a name for the new dataflow.
3. Press OK.
4. Select an Input device in the Available inputs/outputs field
and press the Right arrow button to assign the service to
the dataflow.
5. Repeat step 4 to assign input/output devices as desired.
Most dataflows consist of one input device and one or
several output devices. However in two cases it is possible
to have two input devices, when defining a dataflow with
DICOM Worklist or DICOM Query/Retrieve in combination
with a database.
6. Adjust the device’s parameters as described below.
Query Retrieve Query Retrieve
Retrieve images from a DICOM
server
Dataflow Description