Getting started
40 Vivid i User’s Manual
2378958-100 Rev. 02
To disconnect a probe:
1. Freeze the image by pressing
FREEZE.
2. Press the connector locking lever down to unlock the
connector.
3. Pull the probe and connector straight out of the probe port.
4. Carefully slide the probe and connector away form the
probe port and around the right side of the keyboard.
5. Ensure the cable is free.
6. Be sure that the probe head is clean before placing the
probe in its storage case.
Transporting Probes
When transporting a probe a long distance, store it in its
carrying case.
Storing Probes
It is recommended that all probes be stored in the carrying
case provided.
• First place the probe connector into the carrying case
• Carefully wind the cable into the carrying case.
• Carefully place the probe head into the carrying case.
DO NOT use excessive force or impact the probe face.