7.14.1 Loading system programs
Note!
Always make a backup copy before starting work on
updating system programs.
1. Connect a PC to the machine; see under Connecting a PC.
2. Check that the machine is in STANDBY mode.
3. Make a backup copy by starting CS 1000 and choosing:
Tools/PACS RAM/Upload To File...
Save the *.prm file in your chosen location.
The program will report an error during conversion. Disregard this.
To check that conversion was successful, check the size of the
*.prm file.
The size of the file should be 84416 bytes.
4. If using an existing card, do a Save RAM In Flash (the set
calibration values will then be automatically moved back after
loading of the program(s)).
6. Switch off the power to the machine with the main switch.
7. Move the communication cable to X25.
8. Move jumper 1 on the board for the PACS 300 control system
from Normal to system load.
Check that the three jumpers 2 are in the Flash position and
that battery jumper 3 is in the ON position.
7 REPAIRS AND ADJUSTMENT
TECHNICAL MANUAL GETINGE WD15 CLARO 101 (128)
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>