5.6.2 Checking the filter for incoming water 4.2
In order to ensure a proper supply of water, the valves at the water
connection points must be inspected every year. If this is not done,
the risk of leakage or the tank filling slowly increases, which can
cause operational disturbances. Check the valves on the water
connector by removing and cleaning the flow limiter and filter.
000218
Figure 8. Valve for water connection
5.7 Dosing system (Item 5)
WARNING!
The detergent in the container may be corrosive.
Wear safety glasses and protective gloves.
The following inspection measures are to ensure the proper
operation of the dosing system. Errors may result in inadequate/no
cleaning and/or lime deposits.
1
002482
Figure 9. Container with detergent
1. Empty container alarm
5 PREVENTIVE MAINTENANCE
TECHNICAL MANUAL GETINGE WD15 CLARO 67 (128)