7.3.2 Replacing a dosing pump
Removal
Remove the ramp with the dosing pumps by unscrewing the
two screws that are in the right leg.
Mark the hoses so that the right hose is connected to the right
pump.
Then remove the hoses before the ramp is taken out from its
position.
Assembly
Attach the hoses to each respective pump.
Make sure that the right hose is connected to the right pump.
Put the ramp in and screw it in place with the two screws.
7.3.3 Replacing the waste pump
Removal
Remove the waste pump by unscrewing the two screws that
hold the pump in the right leg.
Take out the pump and loosen the two hoses that are
attached to the pump.
Replace the pump.
Assembly
Attach the two hoses that are connected to the pump.
Make sure that the hose clamps are properly attached.
Screw the pump into place with the two screws.
Check to make sure that no leaks occur.
7.3.4
Replace the pressure switch from the sump and pump pressure
The pressure switch is attached with a hose clamp.
Loosen the clamp and replace the pressure switch.
Make sure that the hose clamps are properly attached.
Check that there are no leaks.