7.3.2 Replacing a dosing pump
Removal
• Remove the ramp with the dosing pumps by unscrewing the
two screws that are in the right leg.
• Mark the hoses so that the right hose is connected to the right
pump.
• Then remove the hoses before the ramp is taken out from its
position.
Assembly
• Attach the hoses to each respective pump.
• Make sure that the right hose is connected to the right pump.
• Put the ramp in and screw it in place with the two screws.
7.3.3 Replacing the waste pump
Removal
• Remove the waste pump by unscrewing the two screws that
hold the pump in the right leg.
• Take out the pump and loosen the two hoses that are
attached to the pump.
• Replace the pump.
Assembly
• Attach the two hoses that are connected to the pump.
• Make sure that the hose clamps are properly attached.
• Screw the pump into place with the two screws.
• Check to make sure that no leaks occur.
7.3.4
Replace the pressure switch from the sump and pump pressure
• The pressure switch is attached with a hose clamp.
• Loosen the clamp and replace the pressure switch.
• Make sure that the hose clamps are properly attached.
• Check that there are no leaks.
7 REPAIRS AND ADJUSTMENT
TECHNICAL MANUAL GETINGE WD15 CLARO 83 (128)
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>