7.14.7 Temperature sensors – with ice bath and oil bath
Proceed as follows to calibrate the temperature sensors with the
aid of ice bath and oil bath:
1. Connect a PC with the CS 1000 program installed for the
disinfector.
2. Prepare an ice bath and an oil bath. The ice bath must
consist of crushed ice in a bowl of cold water. The bath must
stand for at least 20 minutes so that the temperature can
stabilize. The oil bath must be switched on for at least 45
minutes at the set temperature (100ºC) to stabilize the
temperature.
3. Remove the temperature sensors from the disinfector. Tape
the sensors together.
4. Lower the temperature sensors into the ice bath together with
an external thermometer.
5. Start CS 1000 and choose Settings/Calibration/Analog Input
Automatic...
6. Click Temp chamber and Temp chamb indep(A), then click
Calibrate (B).