5.3.2 Checking the cooling fan 1.2
The fan must be functioning properly to prevent internal
components from overheating. If the fan is not working, the
function of various components and their service life can be
affected negatively. Clean the fan blades if there is dirt on them.
The cooling fan is fitted to base plate.
5.3.3
Checking the panel 1.3
The purpose of checking the panel is so that the machine has the
esthetically and hygienically correct appearance and that users
select the correct functions.
1. Check that the panel is not damaged.
2. Check that the display is readable and that the symbols are
clear and legible.
5.3.4 Checking the machine is level 1.4
The purpose of the check is to ensure that the machine does not
become unbalanced and that the door lock is working as intended.
1. Check that the machine is level.
2. Check that the pin on the door is centered in the hole on the
door lock. Adjust the machine's feet as needed.
Figure 5. Pin on the door lock
5.3.5
Checking the Quick guide 1.5
Check that the following are posted close to the machine.
• Quick guide
• Goods positioning sign
5.3.6
Checking error codes and number of processes 1.6
The number of processes and error codes are either read via the
CS1000 service program or via the panel. For more information
see 3 MENU TREE.
5.4
The chamber (Item 2)
5.4.1 Clean the inside of the chamber 2.1
This removes the coating that has formed in the chamber. If it is
not removed, additional dirt and debris can quickly build up in the
chamber. Cleaning is ideally carried out when testing Getinge
Clean Neutralizer Plus.
5 PREVENTIVE MAINTENANCE
TECHNICAL MANUAL GETINGE WD15 CLARO 63 (128)
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>