3.3 A parameters
A-Parameter Default
value
Lower
limit
Upper
limit
Unit Description
TI00 PRE RINSE 1
TIME
00:01:00 00:00:00 00:10:00 - Time Pre-rinse 1
TI01 PRE RINSE 2
TIME
00:01:00 00:00:00 00:10:00 - Time Pre-rinse 2
TI03 WASH INT 1 TIME 00:01:00 00:00:00 00:30:00 - Time Wash 1
LI01 WASHING TEMP 60.0 30.0 90.0 C Temperature Wash 1
LI08 WASHING DOS
TEMP
35.0 20.0 85.0 C Dosing temperature Wash 1
LI23 WASH DOS
AMOUNT
12.0 0.0 50.0 ml Dosing amount Wash 1
TI04 WASH INT 2 TIME 00:01:00 00:00:00 00:30:00 - Time Wash 2
LI03 WASH INT 2
TEMP
60.0 30.0 90.0 C Temperature Wash 2
LI08 WASHING DOS
TEMP
35.0 20.0 85.0 C Dosing temperature Wash
2.
LI33 WASH I 2 DOS
AMO
0.0 0.0 50.0 ml Dosing amount Wash 2
TI05 NEUTR TIME 00:01:00 00:00:00 00:30:00 - Time Neutralization
LI04 NEUTR TEMP 60.0 30.0 90.0 C Temperature Neutralization
LI11 NEUTR DOS
TEMP
35.0 20.0 85.0 C Dosing temperature
Neutralization.
LI38 NEUT DOS
AMOUNT
0.0 0.0 50.0 ml Dosing amount
Neutralization
TI06 RINSE 1 TIME 00:01:00 00:00:00 00:10:00 - Time Rinse 1
TI07 RINSE 2 TIME 00:01:00 00:00:00 00:10:00 - Time Rinse 2
TI08 RINSE 3 TIME 00:01:00 00:00:00 00:10:00 - Time Rinse 3
LI12 F RINSE DOS
TEMP
80.0 20.0 85.0 C Dosing temperature Final
Rinse/Disinfection
LI43 FRINSE DOS
AMOUN
2.0 0.0 50.0 ml Dosing amount Final Rinse/
Disinfection
TI10 DRYING TIME 00:25:00 00:00:00 00:10:00 - Time Ventilation
3 MENU TREE
44 (128) TECHNICAL MANUAL GETINGE WD15 CLARO
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>