7.14.2 Load in a new application into PACS with CS1000
Note: Calibration constants for analog inputs are reloaded
from the flash memory. Calibration constants for dose
monitor get a new value.
1. Make a backup of the existing software. "Tools" "PACS RAM"
"Upload to file (if this has not already been done).
000322
2. Download the new software. “Tools” “PACS RAM” “Download
From File”.
000239
000240
3. Start the machine and check that everything is as it should
be.
4. Switch off the power to the machine and install the “Flash
prom” (if it is missing).
7 REPAIRS AND ADJUSTMENT
104 (128) TECHNICAL MANUAL GETINGE WD15 CLARO
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>