5.7.1 Checking hoses 5.1
The transport hoses can harden and become brittle depending on
which chemicals are used.
Check and replace the hoses on a regular basis to avoid leakages
occurring.
5.7.2
Replacing a hose to a hose pump 5.2
WARNING!
Only to be carried out by authorized staff.
WARNING!
Before starting work, make sure that the machine
is disconnected from the electric power supply.
Remove the cover by unscrewing screw A.
Disconnect the hoses at connections B and C.
Take out the entire unit including hose (a complete spare
parts kit).
Install the new unit.
Refit the cover.
Always check and calibrate the dosing amount after replacing
a pump and/or a hose.
B C
A
000906
5 PREVENTIVE MAINTENANCE
68 (128) TECHNICAL MANUAL GETINGE WD15 CLARO