5.7.1 Checking hoses 5.1
The transport hoses can harden and become brittle depending on
which chemicals are used.
Check and replace the hoses on a regular basis to avoid leakages
occurring.
5.7.2
Replacing a hose to a hose pump 5.2
WARNING!
Only to be carried out by authorized staff.
WARNING!
Before starting work, make sure that the machine
is disconnected from the electric power supply.
• Remove the cover by unscrewing screw A.
• Disconnect the hoses at connections B and C.
• Take out the entire unit including hose (a complete spare
parts kit).
• Install the new unit.
• Refit the cover.
• Always check and calibrate the dosing amount after replacing
a pump and/or a hose.
5 PREVENTIVE MAINTENANCE
68 (128) TECHNICAL MANUAL GETINGE WD15 CLARO
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>