3.4 P parameters
P Parameter Default
value
Lower
limit
Upper
limit
Unit Description
CT00 PRINSE1
FILLTYPE
0 0 3 - Water type Pre-rinse 1
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT01 PRINSE2
FILLTYPE
0 0 3 - Water type Pre-rinse 2
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT02 WASH 1 FILLTYPE 2 0 3 - Water type Washing
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT03 WASHINT
FILLTYPE
2 0 3 - Water type Intensive wash
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT04 NEUTR FILLTYPE 2 0 3 - Water type Neutralization.
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT05 RINSE 1 FILLTYPE 3 0 3 - Water type post-rinse 1
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT06 RINSE 2 FILLTYPE 3 0 3 - Water type postrinse 2
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT07 RINSE 3 FILLTYPE
3 0 3 - Water type postrinse 3
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT08 F RINSE FILLTYPE 3 0 3 - Water type final rinse
0=Cold water, 1=Hot water,
2=Cold+Hot water
3=De-ionized water
CT09 WASH 1 DOS
PUMP
1 0 4 - Dosing pump Wash
0=No dosing
1=Pump 1
2=Pump 2
3=Pump 3
4=Pump 4
3 MENU TREE
TECHNICAL MANUAL GETINGE WD15 CLARO 45 (128)
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>