EasyManuals Logo

Getinge GE15 User Manual

Getinge GE15
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
4.2.3 Emergency opening of the machine
WARNING!
Goods in the machine MUST be reprocessed in
the appropriate process after an Emergency
Opening is performed.
WARNING!
If the goods are still hot after a program is
completed, handle them with care to avoid burns.
In case of power failure, you can open the machine using the cord
attached at the right, inside the lower front door.
Remove the screws.
Unhook the door and fold it downwards.
Pull the cord and the door opens.
1
002236
Figure 4. Cord for emergency opening
1. Door that is folded downwards. There is a emergency
opening cord behind the door.
4 ALARM MANAGEMENT
54 (128) TECHNICAL MANUAL GETINGE WD15 CLARO
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>

Table of Contents

Other manuals for Getinge GE15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge GE15 and is the answer not in the manual?

Getinge GE15 Specifications

General IconGeneral
BrandGetinge
ModelGE15
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals