7.2.2 Replacing/Cleaning strainers
Removal
• Take out the fine strainer from the middle of the plastic inset.
• Then twist a quarter turn counterclockwise around the coarse
strainer.
• Clean or replace the strainers as needed.
•
Figure 13. Fine strainer in bottom of chamber
Assembly
• Replace the coarse strainer.
• Put the plastic insert back and twist it a quarter of a turn
clockwise.
• Put the fine strainer in the middle of the plastic insert.
7.2.3
Replacing/Cleaning spray arms
Clean the spray arms when necessary by unscrewing the central
nut, lifting off the spray arm and rinsing it until it is clean. Take off
the plastic end pieces by pressing the metal bar over the edge of
the spray arm and removing the plastic ends. Using a washing
brush, clean the inside of the spray arm.
Figure 14. Spray arm
After cleaning, press the metal clip back into the groove in the
plastic end piece. Check that the arm can rotate freely when re-
installed.
7 REPAIRS AND ADJUSTMENT
80 (128) TECHNICAL MANUAL GETINGE WD15 CLARO
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>