When the pump stops, read off the amount of detergent remaining
in the beaker and enter the amount used in the “Measured volume”
field (4). Then press Calc. (5).
The calibration constant is now re-calculated and calibration is
complete. If something goes wrong during calibration or if you want
to cancel calibration for some reason, press the ABORT button (6).
Note: The four first dose monitors (DM00-03) run only by
time. If the flow monitors are activated, then the
corresponding dose monitors (DM04-07) that run on pulse
sensors must also be calibrated.