5-18
10
Press [2] key (SCAN). The following message is displayed.
Following message is displayed.
11
Conrm that the tool is at the position shown in the left image
(grey square area), and press the [ENTER] key.
It will scan the registration mark, and then plot the registration mark for
comparison. It will return to SENSOR OFFSET ADJ. screen once the
plotting is completed.
Feed direction of media
Distance adjustment value Y
Distance
adjustment
value X
Registration mark for
adjustment
Registration mark for
comparison
Feed direction of media
Align the tool to this area
Supplement
Position of the registration mark is measured at
the center of the line.
12
Using the registration mark plotted in step 9, measure the
distance of how much the adjustment registration mark
needs to be moved so both will overlap, and record the
value. As an example, in the gure above, it needs to move
in negative directions for both X and Y directions, so both of
the adjustment values will be negative values.
13
Press the [3] key (X).
SENSOR OFFSET ADJ. X screen is displayed.
Supplement
You can set the range between -3.0mm and
+3.0mm.
14
Press the POSITION ( ) keys and increase or decrease the
setting value. Set the X value measured in step 12.
15
Conrm the setting and press the [ESC] key (PREVIOUS).
SENSOR OFFSET ADJ.will be conrmed and it will return to SENSOR
OFFSET ADJ. screen.