Signal Exchange with Machine Control In semi-automatic operation
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
121
Minimal operation in a time-way diagram
ON
OFF
Binary
Program -
number
Main switch OFF/ON
Input: Automatic
Output: Ready for operation
Input: 1
Input: 2
Input: 4
Input: 8
Input: 16
Input: 32
Input: 64
Input: --
Input: Machine ready
Input: Start
Output: No Leak
Output: Rework 1
Output: Leak
Output: Fault
Input: Leak simulation
Input: Static Test (Flooding) On
(Program number 7 is displayed)
The input “Start” should be a pulse with a length of approx. 1 seconds.
* Entry point after removing the dignal “Abort”. This signal deletes all outputs and interrupts the measuring
process at any point (only in automatic mode). The specimen is vented (e.g. for emergency stop).
Measurement with
Leak simulation
Measurement without
Leak simulation
Designation
Flooding