Command reference Menu 3. SETUP
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
Vent:
Enter the vent time in seconds from 0.10 … 999.99.
Default value is 1.00 sec.
NOTE!
The entered vent time influences the output time for the result mes-
sage after completing the measurement.
Vent ON/OFF
You can switch vent to ON or OFF with the – and + keys.
Default setting is OFF.
After these inputs and pressing the SHIFT + PgDn keys, or SHIFT + PgUp keys, the
second page of the dialog SET PARAMETER is displayed:
Continuation of dialog window SET PARAMETER
Unit:
The unit for the measuring value can be selected with the – and + keys: Pascal, mbar,
psi, mmWC, mmHg or ml/min.
Default setting is Pascal.
Fixture offset:
This value influences the zero point of the device. Depending on the direction of the
entered value, the zero point is reduced in case of negative values and increased in
case of positive values.
Enter the fixture offset from 0 … 999 Pa.
Default value is 0 Pa.