Menu 3. SETUP Command reference
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
NOTE!
Change the default values for underfill time, pause time and tare time
only after consulting the manufacturer.
Underfill time: 0.3 s, Pause time: 1.0 s and Tare time: 0.5 s.
Set the times so, that an open specimen connection will cause a rap-
id shut-down.
The underfill time must be in the range from 0.00 … 10.0 seconds.
It is within the stabilizing time and has no influence on the duration of the test cycle.
Default value is 0.30 sec.
Test pressure:
ATTENTION!
With a pressure/vacuum unit do not enter a fill/test pressure of –
0.09 … 0.19 bar/ü.
The value should be 0.2 … 6.00 bar.
Default value is 0.90 bar.
Pressure tolerance:
The value should be 0.5 … 10.0 % and it is the tolerance of the pressure regulation in
% of the end value.
Filling:
Enter the fill time in seconds from 0.10 … 999.99 seconds.
Default value is 3.00 sec.
Stabilize:
Enter the stabilize time in seconds from 0.10 … 999.99.
Default value is 3.00 sec.
NOTE!
Change the values for underfill time, pause time and tare time only
after consulting the manufacturer. Accept the default values for the
underfill time: 0.3 s, Pause time: 1.0 s and Tare time: 0.5 s.
In case of low test pressures (approx. <0.5 bar) or low volume spec-
imen (approx. <100 ml) the pause and tare times may need to be
shortened (possibly no test pressure in specimen).
The times must be set in such a way, that an open specimen will trig-
ger a rapid shut-down.
Pause:
Enter the pause time in seconds from 0.10 … 999.99.
Default value is 1.00 sec.
Auto Zero:
Enter the auto zero time in seconds from 0.10 … 999.99.
Default value is 0.50 sec.
Measuring:
Enter the measure time in seconds from 0.10 … 999.99.
Default value is 3.00 sec.