Command reference Menu 3. SETUP
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
Error number:
The two-digit numerical value specifies the type of the error of the executed test pro-
gram that has occurred:
Pressure switch S2 defective
Pressure switch S3 defective
No pressure in test system
Coarse leak in reference volume
Pressure after filling valve
Temperature value too low
Overflow absolute pressure
transducer
Limit contact
"Leak" cannot be reached
Pressure increases after shutting
off
Differential pressure switch not off
Differential pressure switch not on
Pressure system not adjustable
Rated pressure out of range
Hook 1 missing (optional)
Hook 2 missing (optional)
Hooks 1 +2 missing (optional)
Pressure switch S4 defective
Pressure correction value too high
Pressure in reference volume to
low or flow-in time too short
Pressure in reference volume too
high
Test pressure in sequence too low
Test pressure in sequence too
high
The output of some of the error numbers depends on device type or options. If the
number of the error is <> 0, minus symbols will be displayed for measuring value and
measuring result.
If a numerical value cannot be displayed in the defined output format, the field will be
output with "" (Code 18H) when exceeded, or with "" (Code 19H)
when fallen short of. The individual positions are separated by semicolon. The data set
is closed with a carriage return and a line feed (CR+LF).
Examples:
No Leak 00000001;01;01022000;152621;000000;01;01;00
Gross leak 00000002;01;01022000;152637;------;01;04;00
Fault 00000003;01;01022000;152718;------;01; --;25