Menu 3. SETUP Command reference
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
Serial number:
The serial number consists of 8 digits and starts with 00000001 when the device is
switched on. Once the counter has reached 99999999, it will automatically restart at
00000001.
Program number:
The number of the executed test program in the range from 01 … 99.
Day Month Year:
The date of testing consists of 8 digits: 2 digits for the day, 2 digits for the months, 4
digits for the year.
Hour Minute Second:
The time of testing consists of 6 digits. 2 digits each for hour, minute and second.
Measuring value:
The measured value with floating point representation matches the chosen unit. The
output width is always 6 digits. Error messages or the measuring result "Coarse Leak"
are represented by 6 minus dashes.
Unit:
The two-digit numerical value specifies the unit of the executed test program:
01 = Pa / ml/min
02 = mbar / ml/h
03 = PSI / l/min
04 = mmWC / l/h
05 = mmHg / m³/min
06 = ml/min / m³/h or %
Result:
The two-digit numerical value specifies the measuring result of the executed test program:
01 = I.O. / I.O.
02 = Rework 1 / LT
03 = Leak / UT
06 = Leak / Coarse Leak (Major leak)
-- = Error messages are identified by 2 minus dashes.