14
_____________________________________________________________________
2.5 Connecting the 9322 DIFFERENTIAL PROBE
______________________________________________________________
Grabber cli
Alli
ator cli
Power jack
(Connect the 9418-10
AC ADAPTER
BNC connecto
(Connect to the analog
in
ut terminal
The Model 9322 is a differential probe that connects to
input modules for the 8800 MEMORY HiCORDER series.
High voltage applied between the H and L terminals is
divided by 1/1000 and output as a voltage signal relative to
GND.
Connection Method
1. Connect the alligator or grabber clips to the 9322.
For the 9322, the maximum rated voltage to earth is as
follows.
When using grabber clips: 1500 V AC or DC (CAT ll) /
600 V AC or DC (CAT lll)
When using alligator clips: 1000 V AC or DC (CAT lI) /
600 V AC or DC (CAT lll)
2. Connect the AC ADAPTER to power jack of the 9322.
The indicator lights when power is applied.