69
_____________________________________________________________________
4.9 Numerical Calculation (Measurement)
______________________________________________________________
NOTE
The calculation area can be specified by the A/B cursors. If only
cursor A is used, the calculation area extends from the cursor to
the end of the measurement data. The (horizontal) line cursor is
not used to specify calculation area.
When both numerical calculation and sequential save are
enabled at the same time, calculation starts after sequential save
finishes. All blocks are processed by the calculation, so the
calculation result is refreshed and the final calculation result is
that of the last block. To view the results of all calculations,
enable printing or saving of the calculation results.
Numerical Calculation with New Data
From the Waveform Screen, press the
START
button to
begin measurement. Data input begins when trigger
conditions are met. After the specified data record amount
is input, numerical calculation starts, and when calculation
is finished the results are displayed. If printing or saving of
calculation results has been enabled, those processes
execute.
When starting to input subsequent data, the previous data is
cleared.