9.7 Message Reference
133
9
(6) System
Self-Calibration State and Setting
Execute Self-Calibration
Syntax
Command
:SYSTem:CALibration
Description
Executes self-calibration.
Note
If this command is received while measuring, self-calibration executes after measurement is
finished.
Set and Query Self-Calibration Execution State
Syntax
Command
:SYSTem:CALibration:AUTO <1/ 0/ ON/ OFF>
Query
Response
:SYSTem:CALibration:AUTO?
<ON/ OFF>
<ON> = AUTO Self-Calibration selected
<OFF> = MANUAL Self-Calibration selected
Example
:SYST:CAL:AUTO OFF
:SYST:CAL:AUTO?
OFF
Note
Even when AUTO is selected, Self-Calibration can be manually performed at any time by sending
the
:SYSTem:CALibration command.
Set and Query the Key Beeper Setting
Syntax
Command
:SYSTem:BEEPer:STATe <1/ 0/ ON/ OFF>
Query
Response
:SYSTem:BEEPer:STATe?
<ON/ OFF>
Example
:SYST:BEEP:STAT ON
:SYST:BEEP:STAT?
ON
Select and Query the Line Frequency Setting
Syntax
Command
:SYSTem:LFRequency <AUTO/ 50/ 60>
Query
Response
:SYSTem:LFRequency?
<AUTO/ 50/ 60>
Example
:SYST:LFR 50
:SYST:LFR?
50