9.7 Message Reference
134
Set and Query the Key-Lock State
Syntax
Command
:SYSTem:KLOCk <1/ 0/ ON/ OFF>
Query
Response
:SYSTem:KLOCk?
<ON/ OFF>
Description
Sets and queries the full key-lock state (panel key access to all settings, including comparator
settings, is disabled).
Example
:SYST:KLOC ON
:SYST:KLOC?
OFF
Set and Query the Header Present Setting
Syntax
Command
Query
Response
:SYSTem:HEADer <1/ 0/ ON/ OFF>
:SYSTem:HEADer?
<ON/ OFF>
Example
:SYST:HEAD ON
:SYST:HEAD?
:SYSTEM:HEADER ON
Note
When turning the power on and after the RST command, this is initialized to OFF (no header).
Select the EOM Output Setting
Two modes of EOM (external I/O End-of-Measurement signal) output are available (EOM is asserted at the end of mea-
surement, and de-asserted according to the selected output mode).
See: "8.2 Timing Chart" (p. 88)
EOM Output Mode Setting
Syntax
Command
:SYSTem:EOM:MODE <HOLD/PULSe>
Query
Response
:SYSTem:EOM:MODE?
<HOLD/ PULSE>
<HOLD> = Holds the EOM
signal until measurement starts by the next trigger signal.
<PULSE> = Sets EOM=OFF according to the specified pulse width.
Example
:SYST:EOM:MODE PULS
EOM Pulse Width Setting
Syntax
Command
:SYSTem:EOM:PULSe <Pulse width>
Query
Response
:SYSTem:EOM:PULSe?
<Pulse width>
<Pulse width> = 0.001 to 0.100 (NR2) (seconds)
Example
:SYST:EOM:PULS 0.005