9.7 Message Reference
141
9
Time to receive measured values is different for the :FETCh? and:READ? commands.
See: "9.8 Data Exporting Methods" (p. 148)
Reading the Most Recent Measurement
Syntax
Query
:FETCh?
Description
Reads the most recent measurement. No trigger occurs.
See: "Measurement Value Formats" (p. 140)
Example
:FETC?
1023.579E-3
Note
Binary data is returned when the Binary sending format is selected.
Measuring (Awaiting Triggers and Reading Measurements)
Syntax
Query
:READ?
Description
Switches from the Idle State to the Trigger Wait State, then reads the next measured value. With
auto-ranging enabled, the most suitable range is selected before measurement.
See: "Measurement Value Formats" (p. 140)
Error
This command causes an execution error if issued during the Continuous Measurement state
(after :INITIATE:CONTINUOUS ON).
This command causes an execution error if issued during the Trigger Wait State.
Note
The next command does not execute until measurement is finished.
When using an external trigger source (EXT) with auto-sending of measured values (Auto-
Memory function) enabled, returned measured values will be duplicated. Therefore, disable the
Auto-Memory function when triggering externally.
See: "6.4 Auto-Exporting Measured Values (at End of Measurement) (Data Output Function)" (p.
77)
Binary data is returned when the Binary sending format is selected.
Trigger Source Operation
IMMediate Triggers and reads measured value.
EXTernal Triggers by TRIG
signal (EXT I/O) input, and continuously reads
measured values.