9.7 Message Reference
142
(9) Memory Functions
You can save and load up to 30000 measurement data entries.
Set and Query Memory Functions
Memory Function Execution Mode
Syntax
Command
:MEMory:MODE <OFF/ MEMory/ AUTO>
Query
Response
:MEMory:MODE?
<OFF/ MEMORY/ AUTO>
<OFF> = Memory function OFF
<MEMORY> = Memory function
<AUTO> = Auto memory function
Description
Selects the memory function operation mode.
The Memory function can be disabled (OFF), set to store one data item (MEM), or set for auto-
memory (AUTO) operation.
Example
:MEM:MODE MEM
:MEM:MODE?
AUTO
Note
•
Changing the memory mode setting erases stored data.
• Enabling the auto-memory mode automatically selects internal [INT] triggering and enables sta-
tistical calculation.
Clear Memory Data
Syntax
Command
:MEMory:CLEar
Example
:MEM:CLEA
Retrieve the Memory Data Count
Syntax
Query
Response
:MEMory:COUNt?
<Memory data count >
<Memory data count > = 0 to 30000 (NR1)
Example
:MEM:COUN?
3