9.7 Message Reference
145
9
Set and Query Integration Time (in seconds)
Syntax
Command
:RESistance:APERture <Range>,<Speed>,<Integration time
(seconds)>
Query
Response
:RESistance:APERture? <Range>,<Speed>
<Integration time (seconds)>
Command
:LPResistance:APERture <Range>,<Speed>,<Integration time
(seconds)>
Query
Response
:LPResistance:APERture? <Range>,<Speed>
<Integration time (seconds)>
<Range> = (:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Speed> = SLOW/ MEDium/ FAST
<Integration time> = 0.0001 to 0.1 (seconds)
Description
Sets the integration time, in seconds.
Set and Query Contact Improver Operating Mode
Syntax
Command
:RESistance:CIMProve <Range>,<OFF/ HOLD/ PULSe>
Query
Response
:RESistance:CIMProve? <Range>
<OFF/ HOLD/ PULSE>
Command
:LPResistance:CIMProve <Range>,<OFF/ HOLD/ PULSe>
Query
:LPResistance:CIMProve? <Range>
Response <OFF/ HOLD/ PULSE>
<Range> = (:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
Description
Sets Contact Improver operation to OFF, HOLD or PULSE.
Set and Query Contact Improver Level
Syntax
Command
:RESistance:CIMProve:LEVel <Range>,<Level>
Query
Response
:RESistance:CIMProve:LEVel? <Range>
<Level>
Command
:LPResistance:CIMProve:LEVel <Range>,<Level>
Query
Response
:LPResistance:CIMProve:LEVel? <Range>
<Level>
<Range> = (:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Level> = L1/ L2/ L3/ L4 (applied current)
(L1: 17 mA/ L2: 25 mA/ L3: 35 mA/ L4: 50 mA)
Description
Sets the level of current applied by the Contact Improver function.