3Dimensions System User Guide
Chapter 2: General Information
Page 16 MAN-05085-002 Revision 002
DRAFT Preview Copy-Generated May 30, 2018
Warning:
This system can be dangerous to the patient and the user. Always follow the
safety precautions for x-ray exposures.
Warning:
For exposures except magnification case studies, always use the Face Shield.
Warning:
The Face Shield does not protect the patient from radiation.
Warning:
Risk of entrapment. Make sure that the C-arm has 50 cm (20 inches) of
clearance to any object during C-arm rotation. Do not use Auto Rotation
when C-arm clearance is less than 50 cm (20 inches).
Caution:
The system is a medical device and not a normal computer. Only make approved
changes to the hardware or software. Install this device behind a firewall for network
security. The computer virus protection or network security for this medical device is
not supplied (for example, a computer firewall). The network security and anti-virus
provisions are the responsibility of the user.
Caution:
Do not turn off the Acquisition Workstation circuit breaker except in emergency. The
circuit breaker can turn off the Uninterruptible Power Supply (UPS) and risk data
loss.
Caution:
Risk of data loss. Do not put any magnetic media near or on devices that create any
magnetic fields.
Caution:
Do not use any heat source (like a heating pad) on the image receptor.
Caution:
To prevent possible damage from thermal shock to the Digital Image Receptor,
follow the recommended procedure to turn off the equipment.