2 P/N MAN-01964 Revision 001
DRAFT Preview copy—Generated 7/9/2010
User Manual
Chapter 1—General Information
Safety Information
WARNING! Per North American electrical safety requirements, you
must use a Hospital Grade receptacle to provide a correct
Ground.
WARNING! Do not use the electrical equipment near flammable
anesthetics.
Warning: This device contains dangerous material. Return to Hologic all
material removed from service.
Warning: The user or the service personnel must correct problems before
the system is used.
Warning: The user must arrange for preventive maintenance by an
authorized service representative.
Warning: If a paddle touches possible infectious materials, call your
Infection Control Representative for decontamination
instructions.
Caution: The system is a medical device and not a normal computer. Do not
make changes to the hardware or software that are not authorized.
Install this device behind a firewall for network security. The
computer virus protection or network security for this medical device
is not provided (for example, a computer firewall). The network
security and anti-virus provisions are the responsibility of the user.
Caution: Only use the approved accessories with this equipment. The failure to
follow this caution can cause errors and possible data loss.
Note… Hologic does not provide the Gantry power cable for some countries. If
the power cable is not provided, the installed cable must meet the
following requirements and all local codes that apply: 3 conductor, 8
AWG (10 mm
2
) copper not more than 25 feet (7.62 meters) in length.