P/N MAN-01964 Revision 001 75
DRAFT
Preview copy—Generated 7/9/2010
User Manual
Appendix A—Specifications
Operation and Storage Environment
2.0 Operation and Storage Environment
2.1 General Conditions for Operation
2.2 Storage Environment
2.2.1 Gantry
2.2.2 X-ray Detector
2.2.3 Acquisition Workstation
3.0 Acquisition Workstation Technical Information
Temperature Range
20 °C to 30 °C
Relative Humidity Range
20% to 80% Without condensing moisture
Temperature Range
–10 °C to 40 °C
Relative Humidity Range
(Put in a package for storage in a
building)
0% to 95% Without condensing moisture
Temperature Range
10 °C to 30 °C indefinitely
10 °C to 35°C for a maximum of 12 hours
Maximum rate of temperature change
Less than 10 °C per hour
Relative Humidity Range
(Put in a package for storage in a
building)
10% to 80% Without condensing moisture
Temperature Range
–10 °C to 40 °C
Relative Humidity Range
(Put in a package for storage in a
building)
0% to 95% Without condensing moisture
Operating System Windows XPE
Computer Memory 4 GB RAM minimum
Hard Disk Drive Capacity
700 GB minimum
Storage Device Disks CDRW/CDR/DVD+/-RW/DAT
Display 2 or 3 Megapixels as required
Touchscreen Display 17-inch diagonal, minimum
Network Interface 100/1000Base-T Ethernet
Remote Diagnostics Internet
Graphic User Interface Application specific
Radiation Shield Lead (Pb) equivalent 0.5 mm lead for x-ray energy to 35 kV