6 P/N MAN-01964 Revision 001
DRAFT Preview copy—Generated 7/9/2010
User Manual
Chapter 1—General Information
Safety Information
2.7 Interlocks
The Selenia Dimensions has safety interlocks:
The C-arm vertical drive and rotation is disabled when 45 Newtons (10 pounds) or
greater of compression force is displayed.
If the x-ray button is released before the end of the exposure, the exposure stops and an
alarm message appears.
When in Tomo mode, the system does not allow the Grid in the x-ray field
(Tomosynthesis option).
Mirror and Filter interlocks prevent the x-ray exposure when the Light Field Mirror or the
Filter is not aligned.