NON-GYNECOLOGIC SAMPLE PREPARATION
5.8
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
CAUTION:
Failure to completely pour off the supernatant may produce a sparse sample and an
unsatisfactory slide due to dilution of the cell pellet.
Figure 5-3 Pouring Off Supernatant
After pouring off the supernatant, place the centrifuge tube onto a vortexor and agitate the cell
pellet for 3 seconds. Manual vortexing may be achieved by syringing the pellet back and forth
with a plastic pipette. The intention of this vortexing step is to randomize the cell pellet before
transferring to the PreservCyt
Solution vial and to improve the results of the CytoLyt
Solution
washing procedure.
START