NON-GYNECOLOGIC SAMPLE PREPARATION
5.16
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
Procedure for the Use of DiThioThreitol (DTT) with Mucoid Non-Gyn
Samples
DTT has been shown to be a reagent that is effective in reducing the amount of mucus in respiratory
samples.
1
,
2
DTT Stock Solution
Prepare a stock solution by adding 2.5 g DTT
3
to 30 mL of CytoLyt
Solution.
This solution is suitable for use for 1 week when stored at room temperature (15°C–30°C).
Sample Preparation
This procedure is designed for mucoid non-gyn sample processing. Follow the steps for
processing mucoid specimens on the previous page.
After sample collection (Step 1), but prior to vortexing (Step 2), add 1 mL of the stock DTT
solution to the sample.
Proceed with the remaining sample processing steps as listed.
7. Allow to stand in PreservCyt Solution for 15 minutes (page 5.11)
8. Run on ThinPrep
5000 processor with AutoLoader using
Sequence
Non-Gyn
. Fix, stain, and evaluate.
1. Tockman, MS et al., ‘Safe Separation of Sputum Cells from Mucoid Glycoprotein’ Acta Cytologica 39, 1128
(1995).
2. Tang, C-S, Tang CMC and Kung, TM, ‘Dithiothreitol Homogenization of Prefixed Sputum for Lung Cancer
Detection’, Diagn. Cytopathol. 10, 76 (1994).
3. Available from Amresco, contact a sales representative at 800-448-4442 or
www.amresco-inc.com.