Weekly
Clean the slide cassette area and
AutoLoader slide gripper on the robot arm
Lint-free towel and deionized water
Slide cassette area
Slide cassette area
Clean the sensor and slide holder pneu-
matic cups. Let dry.
Lint-free towel and deionized water
Processing area
Clean around lter plug and
lter puncture areas
Cloth or swab and 70% isopropanol
Filter plug
Filter puncture area
As Needed
Refer to the ThinPrep® 5000 Processor with AutoLoader Operator’s Manual for information on these additional maintenance activities:
• Emptying the waste bottle
• Cleaning the touch screen
Checking and emptying material left in the robot arm elevator shaft
• Cleaning the carousel and the dust cover
©2021 Hologic, Inc. All rights reserved. Refer to the ThinPrep® 5000 Processor Operator’s Manual for full operating
instructions, including any warnings, contraindications and safety information. Contact your local Hologic representa-
tive or, in Canada, call Technical Support at 1-800-442-9892.
• Removing and cleaning the drip trays
• Changing the absorbent pads
• Replacing the HEPA lter
• Replacing the carbon lter
Press the Clean System button to
reposition arms.
AutoLoader slide gripper
2/2
ThinPrep 5000 Processor with AutoLoader Maintenance
Quick Reference Guide
English MAN-03929-001 Rev. 005