1. Unlatch and open the door.
2.Replacethelter.
PayattentiontowhichlteristheHEPAlterandwhichisthecarbonlter.Matchthearrowsforthedirection
ofairow.
3.Closeandlatchthefume
extractor door.
4.Conducttheairowtest.
1/1
ThinPrep 5000 Processor with AutoLoader
Change the HEPA Filter
English MAN-03931-001 Rev. 004
Change the HEPA Filter
ThinPrep® 5000 Processor with AutoLoader