HP 8114A Service Guide
-221
-221
-221
-221
-222
-222
Settings conflict
HP-IB message = “Ext-Width mode only until 20V
amplitude”
Local message = “Ext-Width mode only until 20V
amplitude”
Description:
External width mode is only possible, if the amplitude
is
<
2OV,
because of power problems.
Settings conflict
HP-IB message = “Baseline option only with 50
ohm source
(I
Local message = “Baseline option only
with
50 ohm
source”
Settings conflict
HP-IB message = “No width limits in external
width mode”
Local message = “No width limits in external width
mode”
Settings conflict
HP-IB message = “Double pulse max dutycycle 7
percent if amplitude
>
20V”
Local message = “Double pulse max
dutyc
7
percent if
amp1
>
2OV”
Description:
In Double pulse mode maximum dutycycle is 7
percent, if the amplitude is over
ZOV,
because of
power problems.
Data out of range
HP-IB message = “Parameter out of hard limits”
Local message = “Parameter out of hard limits”
Description:
This is the generic “Parameter out of range” error.
It is issued when more specific information is not
available.
Data out of range
HP-IB message = “Limit reached
-
allowed [range]”
Local message = “Limit reached
-
allowed [range]”
Description:
[range] = range of parameter
Troubleshooting
5-25