HP 8114A Service Guide
Memory error
HP-IB message
=
“Power up state corrupt, set to
*RST state”
Local message = “Power
up
state
corrupl,,
set to
*RST”
Description:
The signature check performed on the instruments
parameter store at power up, failed.
Possible reasons are:
I
The instrument was switched off during a
parameter update
I
The battery used to buffer the instruments RAM
needs replacement
n
The instruments RAM is defective.
Memory error
HP-IB message = “Pattern block corrupt, set to
“RST
state”
Local message = “Pattern corrupt, set to ‘RST”
Description:
The signature check performed on the instruments
pattern block at power up, failed.
Possible reasons are:
w
The instrument was switched off during a
parameter update
H
The battery used to buffer the instruments RAM is
empty
m
The instruments RAM is defective.
SAVe/recall
memory lost
HP-IB message = “[rem/lot] Setting [no] corrupt,”
Local message =
“[rem/lot]
Setting [no] corrupt”
Description:
[rem/lot]
= remote or local, [no] = number input]
Save/Recall memory lost. Indicates that the
nonvolatile data saved
by
the
‘SAV
command has
been lost.
-311
-314
Troubleshooting