HP 8114A Service Guide
-314 SAVeirecall
memory lost
HP-IB message =
“[remiloc]
Setting [no] Firmware
revision nr. incompatible”
Local message =
“[rem/lot]
Setting [no] Firmware
rev. incompat.
”
Description:
(rem/me]
= remote or local,
[no]
= number input]
Each firmware issue has a revision number. The
instrument is trying to read a setting with an
incompatible revision number.
SAVe/recall
memory lost
HP-IB message =
“[rem/lot]
Setting [no] Error
check conflict”
Local message =
“[rem/lot]
Setting [no] Error check
conflict”
Description:
[rem/lot]
= remote or local, [no] = number input]
The error check switch in the specified setting is OFF
but the internal error check is ON. The instrument
cannot recall this setting.
SAVe/recall
memory lost
HP-IB message =
“[rem/lot]
Setting [no] Instrument
configuration incompatible”
Local message =
“[rem/lot]
Setting [no]
Config.
incompatible”
Description:
[rem/lot]
= remote or local, [no] = number input]
Every stored setting contains the instrument
configuration. The stored configuration does not
correspond to the actual instrument configuration.
Selftest
failed
HP-IB message = “Selftest error: Microprocessor
board failed”
Local message = “Selftest error:
Microproc.
board
failed”
Description:
Hardware on the Microprocessorboard does not work.
-314
-314
-330
5.38
Troubleshooting