8. Set the Counter to:
FUNCTION
INPUT A
AUTO TRIG
EXT ARM SELECT
Period A
50
62
OFF
a. Start (ST): leading edge
b. Stop (SP): trailing edge
9.
Select the TRIGGER page on the HP 8114A and set up as follows:
l~~~~~~:~~~~~~~~~~~~i
.a............................................
m
m
1 LIMITS 1 I
CONFIQ
t
The TRIGGER Page Set-up
10.
On the HP 8114A set up the OUTPUT page as shown in the
following illustration:
Configuring the Output Page
11. Check the HP 8114A double pulse delay at the following settings,
pressing
m
to trigger a single cycle each time:
‘Ihble
3-9.
Double Delay Settings and TR Reference
Double Delay
Acceptable
Range
TR Entry
500 ns
474.76
ns
to 625.25
ns
4-4
1
P
949.75
ns
to 1050.25
ps
4-5
5 p”s
4.759 to 5.26
ps
/AS
4-6
60
ps
47.5 to 52.5
ps
ps
4-7
500
ps
475
pLs
to 626
ps
4-8
6
Ill.3
4.75
Ill8
to
5.25
Ins
4-9
50 Ins
47.5
ms
to
52.5
m8
4-10
500
m-3
475
Ills
to 626
ms
4-
11
3.16 Testing the HP g114A