n Set Preset Levels = 50% -50% and press [AUTO LEVEL SET]
n Select the Delta t menu and turn the Time markers On
w
Set START ON EDGE = POSl and STOP ON EDGE = POS2
5.
Press the [PRECISE EDGE FIND) key for each new double delay
setting
6. Check the HP
8114A
double delay at the following settings:
lhble
3-8.
Double Delay Settings and TR Reference
7. Connect the HP 8114A to the Counter as shown:
Connecting HP 8114A to the Counter
Testing the HP
81
MA 3.15