Programming syntax
The information in this section covers syntax for both common commands and SCPI com-
mands. For information not covered here, see the IEEE- 488.2 and SCPI standards.
Command words
Program messages are made up of one or more command words.
Commands and command parameters
Common commands and SCPI commands may or may not use a parameter. The following
are some examples:
*SAV <NRf> Parameter (NRf) required
*RST No parameter used
:INITiate:CONTinuous <b> Parameter <b> required
:SYSTem:PRESet No parameter used.
Put at least one space between the command word and the parameter.
Brackets [ ]: Some command words are enclosed in brackets ([ ]). These brackets are
used to denote an optional command word that does not need to be included in the pro-
gram message. For example:
:INITiate[:IMMediate]
These brackets indicate that :IMMediate is implied (optional) and does not have to used.
Thus, the above command can be sent in one of two ways:
:INTiate or :INTiate:IMMediate
Notice that the optional command is used without the brackets. When using optional
command words in your program, do not include the brackets.
4-32 Remote Operation

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals