:AVERage commands
The :AVERage commands are used to configure and control the filter. The
Filter is explained in Section 3.
:STATe <b>
[:SENSe[1]]:CURRent:AC:AVERage:STATe <b> Control filter for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:STATe <b> Control filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:STATe <b> Control filter for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:STATe <b> Control filter for DCV
[:SENSe[1]]:RESistance:AVERage:STATe <b> Control filter for 2
[:SENSe[1]]:FRESistance:AVERage:STATe <b> Control filter for 4
[:SENSe[1]]:TEMPerature:AVERage:STATe <b> Control filter for TEMP
Parameters <b> = 0 or OFF Disable the digital filter
1 or ON Enable the digital filter
Query :STATe? Query state of digital filter
Description These commands are used to enable or disable the digital averaging filter for
the specified function. When enabled, readings will be filtered according to
how the filter is configured.
:TCONtrol <name>
:SENSe[1]]:CURRent:AC:AVERage:TCONtrol <name> Select filter type for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:TCONtrol <name> Select filter type for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:TCONtrol <name> Select filter type for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:TCONtrol <name> Select filter type for DCV
[:SENSe[1]]:RESistance:AVERage:TCONtrol <name> Select filter type for 2
[:SENSe[1]]:FRESistance:AVERage:TCONtrol <name> Select filter type for 4
[:SENSe[1]]:TEMPerature:AVERage:TCONtrol <name> Select filter type for TEMP
Parameters <name> = REPeat Select repeating filter
MOVing Select moving filter
Query :TCONtrol? Query filter type
Description These commands are used to select the type of averaging filter (REPeat or
MOVing) for the specified function. These filter types are explained in Sec-
tion 3 (see “Filter Modes”).
The number of readings that are averaged by the filter is set with the :AVER-
age:COUNt command. The :AVERage:STATe command is used to enable or
disable the filter. Changing the filter type disables auto filter.
5-46 SCPI Command Reference

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
BrandKeithley
Model2000
CategoryMultimeter
LanguageEnglish

Related product manuals