Page #200 background imageLoading...
Page #200 background image
5-62 SCPI Command Reference
:POSetup <name> command
:POSetup <name>
:SYSTem:POSetup <name> Program power-on defaults
Parameters
<name> = RST Select *RST defaults on power up
PRESet Select :SYSTem:PRESet defaults on power up
SAV0 Select saved defaults on power up
Query
:POSetup? Query power-on setup
Description
This command is used to select the power-on defaults. With RST selected,
the instrument powers up to the *RST default conditions. With PRES select-
ed, the instrument powers up to the :SYStem:PRESet default conditions. De-
fault conditions are listed in the SCPI tables (Tables 5-2 through 5-11).
With the SAV0 parameter selected, the instrument powers-on to the setup
that is saved in the specified location using the *SAV command.
:FRSWitch? command
:FRSWitch?
:SYSTem:FRSWitch? Read INPUTS switch
Description
This query command is used to read the position of the FRONT/REAR IN-
PUTS switch. Switch position code is defined as follows:
1 = Front panel inputs selected
0 = Rear panel inputs selected

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD