:RANGe commands
[:UPPer] <n>
[:SENSe[1]]:CURRent:AC:RANGe[:UPPer] <n> Set measurement range for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe[:UPPer] <n> Set measurement range for DCI
[:SENSe[1]]:VOLTage:AC:RANGe[:UPPer] <n> Set measurement range for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n> Set measurement range for DCV
[:SENSe[1]]:RESistance:RANGe[:UPPer] <n> Set measurement range for 2
[:SENSe[1]]:FRESistance:RANGe[:UPPer] <n> Set measurement range for 4
Parameters <n> = 0 to 3.1 Expected reading is amps (ACI and DCI)
0 to 757.5 Expected reading is AC volts (ACV)
0 to 1010 Expected reading in DC volts (DCV)
0 to 120e6 Expected reading is ohms (
2 and 4)
DEFault 3 (ACI and DCI)
757.5 (ACV)
1000 (DCV)
100e6 and
MINimum 0 (All functions)
MAXimum Same as DEFault
Query :RANGe[:UPPer]? Query ACI measurement range
:RANGe[:UPPer]? DEFault Query *RST default range
:RANGe[:UPPer]? MINimum Query lowest measurement range
:RANGe[:UPPer]? MAXimum Query highest measurement range
Description This command is used to manually select the measurement range for the
specifed measurement function. The range is selected by specifying the ex-
pected reading as an absolute value. The Model 2000 will then go to the most
sensitive range that will accommodate that expected reading. For example,
if you expect a reading of approximately 50mV, simply let the parameter
(<n>) = 0.05 (or 50e-3) in order to select the 100mV range.
SCPI Command Reference 5-41